Setup for the Perten AM 5200-A Moisture Tester

Com Port

The Perten AM 5200-A Moisture Tester has the ability to communicate with the Scale Ticket Interface through the PC’s Com port. The Scale Interface will read in the moisture and test weight values from the Perten AM 5200-A Moisture Tester

Moisture Tester Directly Into The PC

The Moisture Tester can be used directly with the PC by plugging a null modem connector into the back of the PC, then plugging a cable into the null modem and running to the serial to USB converter (one bought only from Perten). The other end of the converter plugs into a com port on the back of the Moisture Tester.

Perten AM 5200-A Setup to Interface with Agvance Grain

From the Perten Main Menu, select General settings.

Next, select Change access right.

Tap in the Access right field.

A key board will display to enter the Administrator password. The General Settings screen will then display.

Choose Local Settings.

Local settings will display and Admin will show in the top bar.

Tap the Down Arrow in the bottom right to go to the next screen.

Ensure the Output Mode says Serial. If not, tap in the field until it says Serial.

The Printer option needs to match the baud rate in the Moisture Tester setting in Agvance. Select Save to finish the Perten AM 5200-A Moisture Tester setup.

Agvance Grain Scale Interface Setup for the Perten AM 5200-A Moisture Tester

  • Model – Select the moisture tester model being used.
  • Com Port – This is the PC Com Port that the cable from the Perten AM 5200-A Moisture Tester is plugged into.
  • Baud Rate – This is the communications speed and must match the baud rate of the Perten AM 5200-A Moisture Tester.
  • Parity – This is used when communicating with the Perten AM 5200-A Moisture Tester and must match the setting of the Perten AM 5200-A Moisture Tester.
  • Data Bits – This is used when communicating with the Perten AM 5200-A Moisture Tester and must match the setting of the Perten AM 5200-A Moisture Tester.
  • Stop Bits – This is used when communicating with the Perten AM 5200-A Moisture Tester and must match the setting of the Perten AM 5200-A Moisture Tester.
  • Moisture Grade Factor – Double-click to select a moisture Grade Factor. This Grade Factor will be used for all Commodities.
  • Test Weight Grade Factor – Double-click to select a test weight Grade Factor. This Grade Factor will be used for all Commodities.

Using the Perten AM 5200-A Moisture Tester on the Scale Ticket

  1. Check the Enable Tester checkbox.
    Note: The Enable Tester checkbox is not available if the Moisture Tester setup is not complete.
  2. When moisture results are returned on the Perten AM 5200-A Moisture Tester, the Capture Values button becomes enabled on the Scale Ticket.
  3. Select the Capture Values button to read the values into the grid.

Ethernet

The Perten AM 5200-A Moisture Tester communicates with the Scale Ticket Interface through an ethernet connection. The Scale Interface receives moisture and test weight values from the Perten AM 5200-A Moisture Tester to import into the Scale Ticket.

If hosted, this requires SSI IS to place security rules on the hosted server in order to use the device.

Moisture Tester Directly Into The PC

The Moisture Tester can be used directly with the PC by connecting the Perten machine to the network with an ethernet connection. With the scale machine on the same network, set up an IP address of the Perten AM 5200-A and the IP address of the scale PC in the Scale Interface and on the Perten settings.

Perten AM 5200-A Setup to Interface with Agvance Grain

Note: An Admin password is needed to make changes to the Perten setup. From the Perten main menu, select System Info.

Scroll through the System Info to the Remote IP Address. Enter the IP address of the Scale Interface’s PC if running Agvance locally or the internal IP of the Terminal Server if Agvance is running in a remote desktop session.

Note: The Remote IP must be accessible inside the Local Area Network or through a VPN. UDP Port 1024 must be opened in the Windows firewall on the Terminal Server.

Continue to scroll down one more screen to enter the Instrument IP which is the Perten’s internal IP address. This is the IP address to re-enter on the Scale Configuration screen for the Moisture Tester.

Finish by selecting Back to get to the Home menu.

Agvance Grain Scale Interface Setup for the Perten AM 5200-A Moisture Tester

Navigate to Hub / Utilities / Scale Interface / Agvance Grain Scale Interface / Setup / Moisture Testers.

  • Model – Select the moisture tester model being used (Perten AM 5200-A).
  • Port – Set to Ethernet.
  • IP Address – Set to the Instrument IP from the Perten System Info.
  • Port Number – Set to the UDP Port from the Perten System Info.
  • Moisture Grade Factor – Double-click to select a Moisture Grade Factor. The Moisture Grade Factor selected will be used for all Commodities.
  • Test Weight Grade Factor – Double-click to select a Test Weight Grade Factor. The Test Weight Grade Factor selected will be used for all Commodities.
  • Import Commodity – If selected, the Commodity Cross Reference will need to be set on the Commodity and selected on the Moisture Tester setup. It is suggested to not set a default Commodity on the scale configuration if this option is used.

Using the Perten AM 5200-A Moisture Tester on the Scale Interface Ticket

C Values button to read the values into the grid.